AGM012  12W Medical wall mount Adapter

 

Features

*Universal AC Input Range

*DOE VI/CoC 5

*Protections: Overcurrent/ Overvoltage/Short Circuit

*Means of Patient Protection Design

*Possess risk analysis report

* UL, cUL60601, EN/IEC60601 with 3rd Edition and CE, FCC approved.

*Warranty two Years

*Application: Medical power, Medical charger(CV mode), Power Supply for Health care

Input

Voltage: 90~264VAC

Frequency: 50~60Hz

Input current: 0.5-0.2A max.

Leakage Current: 0.1mA max. @240VAC /50Hz

Inrush current: 50A @115VAC; 90A @230VAC at 25°C cold start

Output

Part No          Rated O/P(current max.)        Ripple & Noise(mV p-p)

AGM012-S05-z      5-6V/2.4-2/12W                   100

AGM012-S75-z      6-8V/2-1.5A/12W                  100

AGM012-S09-z      8-11V/1.5-1.1A/12W               100

AGM012-S12-z      11-13V/1.1-0.93A/12W             100

AGM012-S15-z      13-16V/0.93-0.75A/12              100

AGM012-S18-z      16-21V/0.75-0.58A/12              100

AGM012-S24-z      21-27V/0.58-0.45A/12W            100

AGM012-S27-z      27-33V/0.45-0.37A/12W            100

AGM012-S48-z      33-48V/0.37-0.25A/12W            100

 

z: US/EU plug

Mechanical Drawing  weight: Appro. 80-100g

 

                                                     

General Specification 

Load Regulation: +/-5% Max.

Line Regulation: +/-1% Max.

Insulation resistance: 50Mohm min. from Input to output

Over Current Hiccup mode; auto-recovery, rate output current >110%

Over Voltage Protection: Protected by Zerner diode, Rated output voltage 110~140%

Short circuit Protection: auto recovery

Setup/Hold up/ Rise time: Typ.: 1000mS/10mS/50mS at full load

Efficiency : DOE VI

Derating : Derated from 100% at +40°C linearly to 50% at +70°C

Temperature Coefficient: All output +/-0.04% / °C max.

Withstand Voltage: 5656VDC from input to output

Operating Temperature: 0~40degree C

Storage Temperature: -40~85degree C

Storage Humidity: 5~95% RH

Transient Response: Max. Excursion of 4% better on all models. Recovering to 1% of final value within

500uS. after a 25% step load change

Safety approvals: UL, cUL60601, EN/IEC60601 with 3rd Edition

EMC:  EN600601-1-2: 2007, FCC Part 15 & Part 18 class B

Safety: UL60601-1: 2005, CSA: C22.2 60601-1: 2008, EN60601-1:2006