AGM018  18W Medical wall mount Adapter

 

Features

*Universal AC Input Range

*DOE V

*Pass LPS

*Protections: Overcurrent/ Overvoltage/Short Circuit

*Means of Patient Protection Design

*Possess risk analysis report

* UL, cUL60601, EN/IEC60601 with 3rd Edition and CE, FCC approved.

*Warranty two Years

*Application: Medical power, Medical charger(CV mode), Power Supply for Health care

Input

Voltage: 90~264VAC

Frequency: 50~60Hz

Input current: 0.45-0.2A max.

Leakage Current: 0.1mA max. @240VAC /50Hz

Inrush current: 50A @115VAC; 90A @230VAC at 25°C cold start

Output

Part No          Rated O/P(current max.)        Ripple & Noise(mV p-p)

AGM018-S05-z      5-6V/3-2.5/15W                   80

AGM018-S75-z      6-8V/2.5-1.87A/15W               80

AGM018-S09-z      8-11V/2.25-1.64A/18W            100

AGM018-S12-z      11-13V/1.64-1.38A/18W           120

AGM018-S15-z      13-16V/1.38-1.13A/18W           150

AGM018-S18-z      16-21V/1.13-0.86A/18W           150

AGM018-S24-z      21-27V/0.86-0.67A/18W           150

AGM018-S27-z      27-33V/0.67-0.55A/18W           240

AGM018-S48-z      33-58V/0.55-0.32A/18W           240

 

z: US/EU plug, z: US/EU plug, 2pin & 3pin AU plug, UK plug

Mechanical Drawing  weight: Appro. 130-140g

 

                                                     

General Specification 

Load Regulation: +/-5% Max.

Line Regulation: +/-1% Max.

Insulation resistance: 50Mohm min. from Input to output

Over Current: auto-recovery

Over Voltage Protection: Protected by Zerner diode

Short circuit Protection: Hiccup mode; auto recovery

Setup/Hold up/ Rise time: Typ.: 1000mS/10mS/50mS at full load

Efficiency : DOE V

Derating : Derated from 100% at +40°C linearly to 50% at +70°C

Temperature Coefficient: All output +/-0.04% / °C max.

Withstand Voltage: 5656VDC from input to output

Operating Temperature: 0~40degree C

Storage Temperature: -40~85degree C

Storage Humidity: 5~95% RH

Transient Response: Max. Excursion of 4% better on all models. Recovering to 1% of final value within

500uS. after a 25% step load change

EMC:  EN600601-1-2: 2007, FCC Part 15 & Part 18 class B

Safety: UL60601-1: 2005, CSA: C22.2 60601-1: 2008, EN60601-1:2006