AGM070  70W Medical Desktop Adapter

 

Features

*Universal AC Input Range

*DOE V

*Low Leakage current

*Protections: Overcurrent/ Overvoltage/Short circuit

*Means of Patient Protection Design

*Possess risk analysis report

* UL, cUL60601, EN/IEC60601 with 3rd Edition and CE, FCC approved.

*Warranty two Years

*Application: Medical power, Medical charger(CV mode), Power Supply for Health care

Input

Voltage: 90~264VAC

Frequency: 50~60Hz

Input current: 1.8-0.8A max.

Leakage Current: <0.3mA

Inrush current: 100A @230VAC at 25°C cold start

Output                                                                                                                                                

Part No          Rated O/P(current max.)        Ripple & Noise(mV p-p)

AGM070-S05-z      5-6V/7.5-6.36A/37.5W          200

AGM070-S75-z      6-8V/7.5-5.63A/45W            100

AGM070-S09-z      8-11V/7.5-5.46A/60W           120

AGM070-S12-z      11-13V/6.37-5.39A/70W         120

AGM070-S15-z      13-16V/5.39-4.38A/70W         120

AGM070-S18-z      16-21V/4.38-3.34A/70W         180

AGM070-S24-z      21-27V/3.34-2.6A/70W          180

AGM070-S27-z      27-33V/2.6-2.13A/70W          240

AGM070-S48-z      33-58V/2.13-1.22A/70W         240

 

z: C6/C14 inlet

Mechanical Drawing  weight: Appro. 370-375g

 

                                                     

General Specification 

Load Regulation: +/-5% Max.

Line Regulation: +/-1% Max.

Insulation resistance: 50Mohm min. from Input to output

Over Current: >120% output current, auto recovery

Over Voltage Protection: >140% output voltage

Short Circuit Protection: Hiccup mode; auto recovery

Hold up time: Typ.:  8mS min.

Derating : Derated from 100% at +40°C linearly to 50% at +70°C

Temperature Coefficient: All output +/-0.04% / °C max.

Withstand Voltage: 5656VDC from input to output

Operating Temperature: 0~40degree C

Storage Temperature: -40~85degree C

Storage Humidity: 5~95% RH

Transient Response: Max. Excursion of 4% better on all models. Recovering to 1% of final value within

500uS. after a 25% step load change

EMC:  EN600601-1-2: 2007, FCC Part 15 & Part 18 class B

Safety: UL60601-1: 2005, CSA: C22.2 60601-1: 2008, EN60601-1:2006